- Illuminance Meter
- model no. : T-10A / 10MA
- Desc. :
KONICA MINOLTA
- Application :
Indoor lighting, sun light, light box for advertisement, street light, tunnel lighting, ANSI lumen of projector, effect of light to animals or plants
- Feature :
To measure light intensity with lux or fcd, wide range, accurate, easy to use, compact, data output to PC, support multi-point measurement
-
-
Model: T-10A (standard receptor); T-10MA (mini receptor) Standard compliance: JIS C 1609-1:2006 class AA Sensor: Silicon photocell Relative spectral response: Within 6% (f1´) of the CIE spectral luminous efficiency V (λ) Cosine Correction Characteristics: Within ±1% at 10°; Within ±2% at 30°; Within ±6% at 50°;
Within ±7% at 60°; Within ±25% at 80°
Illuminance unit: lux (lx) or foot candles (fcd) (switchable) Measurement range: Illuminance : 0.01 ~ 299,900 lx, 0.001 ~ 29,990 fcdIntegrated illuminance : 0.01 ~ 999,900 x klx•h,0.001 ~ 99,990 x kfcd•h / 0.001 ~ 9999 hUser calibration function: CCF (Color Correction Factor) setting function Linearity: Within ±2% ±1 digit of displayed value
(based on Konica Minolta standard)
Temperature / humidity drift: Within ±3% ±1 digit of displayed value
(at 20°C / 68°F within operating temperature / humidity range)
Digital output: USB Display: 3 or 4 significant digits (LCD with back-light illumination < 10 lx) Power source: 2 AA-size batteries or AC adapter (optional) Dimensions: Main body & standard receptor : 69 x 174 x 35 mm;
main body : 69 x 161.5 x 30 mm
mini receptor : Ø16.5 x 13.8 mm; cord length : 1 mWeight: 200 g (without battery); 205 g (without battery)